Features and Functions 4
Keysight 34970A/34972A User’s Guide 171
– The name can contain up to 12 characters. The first character must be a letter
(A-Z), but the remaining 11 characters can be letters, numbers (0-9), or the
underscore character (“ _ ”). Blank spaces are not allowed. An error is
generated if you specify a name with more than 12 characters.
–A Factory Reset (*RST command) does not affect the configurations stored in
memory. Once a state is stored, it remains until it is overwritten or specifically
deleted.
– Front-Panel Operation:
NAME STATE , STORE STATE , RECALL STATE
After recalling a stored state, you will notice that a new choice
(UNDO RECALL) is added under RECALL STATE. This allows you to cancel the
last recall operation and revert to the previous state. You can also select
LAST PWR DOWN to recall the state of the instrument at the power-down.
To configure the instrument to recall the power-down state or issue a Factory
Reset when power is restored, select from the following.
PWR ON LAST , PWR ON RESET
– Remote Interface Operation: Use the following commands to store and recall
instrument states (state “0” is the state of the instrument at power down).
*SAV {0|1|2|3|4|5}
*RCL {0|1|2|3|4|5}
To assign a name to a stored state to be recalled from the front panel, send the
following command. From the remote interface, you can only recall a stored
state using a number (0 through 5).
MEM:STATE:NAME 1,TEST_RACK_1
To configure the instrument to automatically issue a Factory Reset (*RST
command) when power is restored, send the following command.
MEMory:STATe:RECall:AUTO OFF