EasyManua.ls Logo

Keysight 34972A - Page 199

Keysight 34972A
353 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Features and Functions 4
Keysight 34970A/34972A User’s Guide 199
To unsecure for calibration
You can unsecure the instrument either from the front panel or over the remote
interface. The instrument is secured when shipped from the factory and the
security code is set to “KT034970” or “KT034972”, depending on the product
number.
Once you enter a security code, that code must be used for both front-panel
and remote operation. For example, if you secure the instrument from the front
panel, you must use that same code to unsecure it from the remote interface.
Front-Panel Operation:
UNSECURE CAL
When you first enter the Utility menu, the calibration entries toggle between
CAL SECURED and UNSECURE CAL. To unsecure the instrument, select
UNSECURE CAL and press . After entering the correct security code,
press again. When you return to the menu, you will see new choices CAL
UNSECURED and SECURE CAL.
Remote Interface Operation: To unsecure the instrument, send the following
command with the correct security code.
CAL:SECURE:STATE OFF,KT034970
NOTE
If you enter the wrong secure code, NO MATCH is displayed and a new choice,
EXIT, is shown.