Error Messages 5
Keysight 34970A/34972A User’s Guide 239
221 Settings conflict: calculate limit state forced off
If you plan to use scaling on a channel which will also use alarms, be sure to
configure the scaling values first. This error is generated if you attempt to assign
the alarm limits first and the instrument will turn off alarms and clear the limit
values.
222 Settings conflict: module type does not match stored state
Before recalling a stored state, the instrument verifies that the same module
types are installed in each slot. The instrument has detected a different module
type in one or more slots.
223 Settings conflict: trig source changed to IMM
This error is generated if you attempt to set the channel advance source
(ROUTe:CHAN:ADVance:SOURce command) to the same source used for the
scan trigger (TRIGger:SOURce command). The command has been accepted
and executed but the scan trigger source is reset to “IMMediate”.
224 Settings conflict: chan ad v source changed to IMM
This error is generated if you attempt to set the scan trigger source
(TRIGger:SOURce command) to the same source used for the channel advance
source (ROUTe:CHAN:ADVance:SOURce command). The command has been
accepted and executed but the channel advance source is reset to “IMMediate”.
225 Settings conflict: DMM disabled or missing
This command is valid only when the internal DMM is installed and enabled. Use
the INSTrument:DMM? command to determine the state of the internal DMM.
For more information, see “Internal DMM Disable” on page 175.
226 Settings conflict: DMM enabled
When the internal DMM is enabled, the ROUTe:CHAN:ADVance:SOURce and
ROUTe:CHAN:FWIRe commands are not allowed. Use the INSTrument:DMM?
command to determine the state of the internal DMM. For more information, see
“Internal DMM Disable” on page 175.
251 Unsupported temperature transducer type
An invalid RTD or thermistor type has been specified. The following RTDs are
supported: α = 0.00385 (“85”) and α = 0.00391 (“91”). The following thermistors
are supported: 2.2 kΩ (“2252”), 5 kΩ (“5000”), and 10 kΩ (“10000”).
Example: CONF:TEMP RTD,1,(@101)