7Tutorial
320 Keysight 34970A/34972A User’s Guide
bank of channels from the other, effectively removing any parallel adjacent
capacitance from the isolated bank. During a scan, the instrument automatically
controls the bank switches.
If you are not using all of the channels on the multiplexer, divide the input signals
equally between Bank 1 and Bank 2. For example, if you are multiplexing eight
channels using the 16-channel multiplexer, use four channels in the lower bank
and four channels in the upper bank. For even better noise immunity, leave one
unused, grounded channel between each input channel.
Module Bank 1 Bank 2
34901A Channels 1 through 10 Channels 11 through 20
34902Α Channels 1 through 8 Channels 9 through 16
34908A Channels 1 through 20 Channels 21 through 40
Source
Source
Source
Bank 1
Bank 2
Channel
DMM
Bank Switch
Bank Switch
C
adj
C
adj
Source
Switch
Channel
Switch