Quick Start 1
Keysight 34970A/34972A User’s Guide 43
For this example, configure channel 103 to measure a J-type thermocouple
with 0.1 °C of display resolution.
3 Run the scan and store the readings in non-volatile memory.
The instrument automatically scans the configured channels in consecutive
order from slot 100 through slot 300 (the SCAN annunciator turns on).
Channels that are not configured are skipped during the scan. In the default
configuration, the instrument continuously scans the configured channels at a
10-second interval.
Press and hold to stop the scan.
4 View the data from the scan.
All readings taken during a scan are automatically time stamped and stored in
non-volatile memory. During the scan, the instrument calculates and stores
the minimum, maximum, and average on all channels in the scan list. You can
read the contents of memory at any time, even during a scan.
From the front panel, data is available for the last 100 readings on each
channel readings taken during a scan (all of the data is available from the
remote interface). From the View menu, select
READINGS
and press
again. Then press and to choose the data you want to view for the
selected channel as shown in the table below.
NOTE
Press to sequentially step through the scan list and take a measurement on
each channel (readings are not stored in memory). This is an easy way to verify
your wiring connections before initiating the scan.