3 53220A/53230A Measurements
126 Keysight 53220A/53230A User’s Guide
– channel is optional channel 3 specified as (@3).
Carrier frequency measurement examples
// measure the carrier frequency using MEASure? and its
// default settings
*RST // reset to start from known state
MEAS:FREQ:BURS? (@3)
-----------------------------------------------------
// measure the carrier frequency using CONFigure, set
// automatic gate configuration
*RST // reset to start from known state
CONF:FREQ:BURS (@3)
INP3:BURS:LEV -6 // set a detector threshold of -6 dB
SENS:FREQ:BURS:GATE:NARR OFF // narrow pulse mode off
SENS:FREQ:BURS:GATE:AUTO ON // automatic gate set up
READ?
Notes
1 See Burst measurement detector threshold in Chapter 4 for information on
setting the threshold for pulse measurements.
2 See Burst carrier frequency measurements in Chapter 5 for information on
narrow pulse mode and on automatic and manual gate control for carrier
measurements.