EasyManua.ls Logo

Keysight 53230A - Pulse PRI and PRF

Keysight 53230A
303 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
53220A/53230A Measurements 3
Keysight 53220A/53230A User’s Guide 127
Pulse PRI and PRF
The commands used to measure the pulse repetition interval (PRI) and pulse
repetition frequency (PRF) (Figure 3-13) of a burst signal are:
MEASure:FREQuency:PRI? [{<expected> | MINimum | MAXimum | DEFault}
[,{<resolution> | MINimum | MAXimum | DEFault}]] [,<channel>]
CONFigure:FREQuency:PRI [{<expected> | MINimum | MAXimum | DEFault}
[,{<resolution> | MINimum | MAXimum | DEFault}]] [,<channel>]
- expected is the expected interval. resolution is the desired measurement
resolution in seconds. The parameter ranges are:
expected: 100 ns to 1.0s (default = 0.001s)
resolution: 1.0E-15 * expected to 1.0E-5 * expected
(default resolution corresponds to a 0.1s gate time)
channel is counter channel 3 specified as (@3).
MEASure:FREQuency:PRF? [{<expected> | MINimum | MAXimum | DEFault}
[,{<resolution> | MINimum | MAXimum | DEFault}]] [,<channel>]
CONFigure:FREQuency:PRF [{<expected> | MINimum | MAXimum | DEFault}
[,{<resolution> | MINimum | MAXimum | DEFault}]] [,<channel>]
expected is the expected repetition frequency. resolution is the desired
measurement resolution in hertz. The parameter ranges are:
expected: 1 Hz - 10 MHz (default = 1 kHz)
resolution: 1.0E-15 * expected to 1.0E-5 * expected
(default resolution corresponds to a 0.1s gate time)
channel is counter channel 3 specified as (@3).

Table of Contents

Other manuals for Keysight 53230A

Related product manuals