53220A/53230A Measurements 3
Keysight 53220A/53230A User’s Guide 127
Pulse PRI and PRF
The commands used to measure the pulse repetition interval (PRI) and pulse
repetition frequency (PRF) (Figure 3-13) of a burst signal are:
MEASure:FREQuency:PRI? [{<expected> | MINimum | MAXimum | DEFault}
[,{<resolution> | MINimum | MAXimum | DEFault}]] [,<channel>]
CONFigure:FREQuency:PRI [{<expected> | MINimum | MAXimum | DEFault}
[,{<resolution> | MINimum | MAXimum | DEFault}]] [,<channel>]
- expected is the expected interval. resolution is the desired measurement
resolution in seconds. The parameter ranges are:
expected: 100 ns to 1.0s (default = 0.001s)
resolution: 1.0E-15 * expected to 1.0E-5 * expected
(default resolution corresponds to a 0.1s gate time)
– channel is counter channel 3 specified as (@3).
MEASure:FREQuency:PRF? [{<expected> | MINimum | MAXimum | DEFault}
[,{<resolution> | MINimum | MAXimum | DEFault}]] [,<channel>]
CONFigure:FREQuency:PRF [{<expected> | MINimum | MAXimum | DEFault}
[,{<resolution> | MINimum | MAXimum | DEFault}]] [,<channel>]
– expected is the expected repetition frequency. resolution is the desired
measurement resolution in hertz. The parameter ranges are:
expected: 1 Hz - 10 MHz (default = 1 kHz)
resolution: 1.0E-15 * expected to 1.0E-5 * expected
(default resolution corresponds to a 0.1s gate time)
– channel is counter channel 3 specified as (@3).