2 53220A/53230A Software Installation and Interface Configuration
52 Keysight 53220A/53230A User’s Guide
Web interface overview
The following information is an introduction to the counter’s Web interface.
The counter welcome page (Figure 2-1) displays IO information, can be used to
identify a connected instrument, and contains menu tabs for accessing additional
counter functionality.
To easily identify the counter among other instruments using its Web interface
home page, clicking on:
Turn On Front Panel Identification Indicator
changes the counter display to “LXI Web Identify” until the indicator is turned off
by clicking on:
Turn Off Front Panel Identification Indicator
Programming the counter
Selecting the ‘Remote Control’ icon (second from top) on the welcome page opens
the Interactive IO window shown in Figure 2-2. From this window, SCPI
commands can be sent to the instrument.
NOTE
Web interface pages other than the ‘Welcome Page’ may be password
protected. When shipped from the factory no password is set; however, an ‘Enter
Password’ dialog box may appear. Click on the dialog box to continue.
If the page is password protected and the password is unknown, pressing
[Utility], (I/O Config), (LAN Reset) on the front panel clears the password.