Triggering and Gating 5
Keysight 53220A/53230A User’s Guide 201
Gate hold off and stop configuration example
The following example shows the sequence commonly used when configuring the
stop gate hold off and stop gate using low-level commands.
//configure the counter for an externally gated time
//interval measurement. The gate start and stop signals are
//applied to the rear panel ‘Gate In’ BNC. Gate closing is
//held off for 1 second after the stop gate signal is
//received.
//
CONF:TINT (@1),(@2) // time interval meas ch1-ch2
TRIG:SOUR INT // use internal trigger source
TRIG:COUN 1 // specify one trigger
SAMP:COUN 2 // set 2 readings (gate cycles)
SENS:GATE:STAR:SOUR EXT // set an external gate source
SENS:GATE:EXT:SOUR EXT // select rear panel Gate In BNC
SENS:GATE:STAR:SLOP POS // set a pos. start gate slope
OUTP:STAT OFF // disable BNC as an output
//
// configure hold off and stop gate
//
SENS:GATE:STOP:HOLD:SOUR TIME // hold off source
SENS:GATE:STOP:HOLD:TIME 1 //hold off gate close for 1s
SENS:GATE:STOP:SOUR EXT // external stop gate source
SENS:GATE:EXT:SOUR EXT // select rear panel Gate In BNC
SENS:GATE:STOP:SLOP POS // select pos. stop gate source
SENS:TINT:GATE SOUR ADV // low-level gate control
READ? // initiate counter and take reading