53220A/53230A Math, Graphing, and Data Logging 6
Keysight 53220A/53230A User’s Guide 231
an instrument reset or present - *RST, SYSTem:PRESet
These actions also clear all readings in reading memory and in the output buffer.
To clear histogram data without clearing the readings in reading memory, the
following command is used:
CALCulate2:TRANsform:HISTogram:CLEar[:IMMediate]
Saving readings
The readings used to create the histogram can be saved to internal flash memory
or to an external USB memory device as comma-separated values (CSV) in one
measurement per line ASCII format.
Selecting Save Readings opens the Export action window from which a path and file
name can be selected or created. The number of readings exported is the total
number in reading memory at the time Export Readings is pressed - which may not
be the total reading count (TRIGger:COUNt x SAMPle:COUNt).
Exporting readings clears reading memory and re-starts the histogram on the
next trigger received.