53220A/53230A Measurements 3
Keysight 53220A/53230A User’s Guide 79
INTernal selects the counter’s internal 10 MHz oscillator. The signal is a
0.5 Vrms (into 50 ) sine wave. The internal oscillator signal is also present on
the counter’s rear panel Int Ref Out connector.
EXTernal selects an external reference signal applied to the rear panel Ext Ref
In connector. The signal must be:
1 MHz , 5 MHz, or 10 MHz
100 mVrms to 2.5 Vrms
–sine wave
and the frequency must be specified by the
SENSe:ROSCillator:EXTernal:FREQuency command.
:AUTO ON enables automatic selection of the reference oscillator source. If a 1,
5, or 10 MHz signal is present on the counter’s Ext Ref In connector, source
EXTernal is set. If a valid signal is not present or is lost, the source
automatically switches to INTernal.
Note that specifying an oscillator source (INTernal or EXTernal) using
[SENSe:]ROSCillator:SOURce disables automatic selection.
:AUTO Off disables automatic selection of the oscillator source. The source is
then set by the [SENSe:]ROSCillator:SOURce command.
The reference oscillator source is set to INTernal with automatic selection
enabled (On) when the counter is shipped from the factory or following the
SYSTem:SECure:IMMediate command. Settings are stored in non-volatile
memory and are not changed after a reset (*RST) or instrument preset
(SYSTem:PREset or Preset key).