106 Keysight E5092A ENA Series Configurable Multiport Test Set
Performance Tests
E5092A Performance Test Record
2-
Insertion Loss
Port 4 (4C) 50M to 1.3G < -16.0 0.30
1.3G to 4G < -11.0 0.23
4G to 8G < -8.0 0.12
8G to 10G < -7.0 0.13
10G to 20G < -4.0 0.10
Port 4 (4D) 50M to 1.3G < -16.0 0.30
1.3G to 4G < -11.0 0.23
4G to 8G < -8.0 0.12
8G to 10G < -7.0 0.13
10G to 20G < -4.0 0.10
From Port -
To Port
Frequency
Range [Hz]
Test Limit
[dB]
Test Result [dB] Measurement
Uncertainty [dB]
Port 4- 4A 50M to 100M > -4.0 0.27
100M to 2G > -3.5 0.24
2G to 3G > -4.5 0.18
3G to 4G > -5.5 0.19
4G to 6G > -6.0 0.11
6G to 8G > -7.5 0.09
8G to 10G > -8.5 0.24
10G to 14G > -9.5 0.29
14G to 18G > -10.5 0.29
18G to 20G > -12.0 0.63
Port
(Selected
Port)
Frequency
Range [Hz]
Test Limit
[dB]
Test Result [dB] Measurement
Uncertainty [dB]