EasyManua.ls Logo

Keysight ENA Series - SP4 T Raptor Holder Assembly Removal

Keysight ENA Series
217 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
198 Keysight E5092A ENA Series Configurable Multiport Test Set
Replacement Procedure
Replacement Procedure for E5092A-020
5-
SP4T Raptor Holder Assembly Removal
Tools Required
TORX screwdriver, T10, T15 and T20.
Box wrench 5.5mm
Torque wrench, 5/8 in.
Procedure
Refer to Figure 5-9 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 189.
Step 2. Detach the front chassis as described in “Front Chassis Detachment” on
page 190.
Step 3. Remove the front frame as described in “Front Frame Removal” on page 191.
Step 4. Remove the assembly board as described in “Main Board Removal” on
page 195.
Step 5. Remove the SP4T switch assembly as described in “SP4T Switch Assembly
Removal” on page 196.
Step 6. Remove the two TORX T10 screw (item 1) on the SP4T raptor holder assembly
and remove the cables.
Step 7. Remove the four TORX T10 screw (item 2) holding the SP4T raptor holder
assembly and lift the raptor holder assembly.
Step 8. Remove the semirigid cable (item 3) on the SPDT switch assembly.
Step 9. Remove the six TORX T10 screw (item 4) holding the SP4T raptor holder
assembly and lift the raptor holder assembly.

Table of Contents