18 Keysight E5092A ENA Series Configurable Multiport Test Set
Performance Tests
2. DC SOURCE MAXIMUM CURRENT TEST
2-
2. DC SOURCE MAXIMUM CURRENT TEST
Description
This test measures limited current value for each control line.
Specification
@ 23 °C 5 °C (Warming up over 90 minutes)
150 mA for each source.
Test equipment
DC Power Supply Keysight E3631A
Digital Multimeter Keysight 34410A
Multifunction Switch/Measure Unit Keysight 34980A
40-Channel Armature Multiplexer for 34980A Keysight 34921A
ET Keysight E5092-66590