108 Keysight E5092A ENA Series Configurable Multiport Test Set
Performance Tests
E5092A Performance Test Record
2-
Isolation
From Common Port To Unselected Port
Port 4 - 4D 50M to 100M > -4.0 0.27
100M to 2G > -3.5 0.24
2G to 3G > -4.5 0.18
3G to 4G > -5.5 0.19
4G to 6G > -6.0 0.11
6G to 8G > -7.5 0.09
8G to 10G > -8.5 0.24
10G to 14G > -9.5 0.29
14G to 18G > -10.5 0.29
18G to 20G > -12.0 0.63
Port - To Port
(Selected Port)
Frequency
Range [Hz]
Test Limit
[dB]
Test Result [dB] Measurement
Uncertainty
[dB]
Port 4 - 4A
(4B)
50M to 500M < -65.0 0.35
500M to 1G < -80.0 0.48
1G to 2G < -85.0 0.60
2G to 6G < -90.0 0.76
6G to 10G < -85.0 0.66
10G to 18G < -75.0 0.54
18G to 20G < -65.0 0.43
Port 4 - 4A
(4C)
50M to 500M < -65.0 0.35
500M to 1G < -80.0 0.48
1G to 2G < -85.0 0.60
2G to 6G < -90.0 0.76
6G to 10G < -85.0 0.66
10G to 18G < -75.0 0.54
18G to 20G < -65.0 0.43
From Port -
To Port
Frequency
Range [Hz]
Test Limit
[dB]
Test Result [dB] Measurement
Uncertainty [dB]