Keysight E5092A ENA Series Configurable Multiport Test Set 149
Troubleshooting
Power Supply Troubleshooting
• If the voltmeter reading is not within the following limits, replace the power
supply unit.
• If the voltmeter reading is within the limits, the power supply is verified.
Table 3-1 Power Supply Unit Output
Supply Range
+ 5 V + 4.75 to + 5.25
+ 15 V + 14.25 to 15.75
- 15 V - 14.25 to -15.75