51
Keysight E5092A ENA Series Configurable Multiport Test Set
Service Guide
Load Match, Insertion Loss and Isolation Test for Switch 1 (SP4T, Port 1, 1A, 1B,
1C and 1D)
Load Match (Unselected)
Port
(Selected
Port)
Frequency
Range [Hz]
Test Limit
[dB]
Test Result [dB] Measurement
Uncertainty [dB]
1B (1A) 50M to 3G < -17.0 0.43
3G to 10G < -11.0 0.37
10G to 16G < -8.0 0.28
16G to 18G < -6.0 0.28
18G to 20G < -4.0 0.35
1C (1A) 50M to 3G < -17.0 0.43
3G to 10G < -11.0 0.37
10G to 16G < -8.0 0.28
16G to 18G < -6.0 0.28
18G to 20G < -4.0 0.35
1D (1A) 50M to 3G < -17.0 0.43
3G to 10G < -11.0 0.37
10G to 16G < -8.0 0.28
16G to 18G < -6.0 0.28
18G to 20G < -4.0 0.35
1A (1B) 50M to 3G < -17.0 0.43
3G to 10G < -11.0 0.37
10G to 16G < -8.0 0.28
16G to 18G < -6.0 0.28
18G to 20G < -4.0 0.35