LACO ORIGINAL OPERATING INSTRUCTIONS TITANTEST
80
© 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807
16. If use use an external backing pump, re-establish the KF 25 connection
to the external backing pump.
17. Re-connect the power supply.
18. Enter your maintenance work on the TMP in " Setup > Maintenance &
Service > Maintenance components".
This is only possible if maintenance is enabled in " Setup > Access con-
trol > Maintenance enabled".
8.6.1.2 Replace SplitFlow 80 for bearing change
Recommended bearing change at least every 4 years, with high loading ear-
lier where necessary. A bearing change can only be carried out by LACO em-
ployees.
8.6.2 Valve block
Maintenance work for:
TITANTEST™ Wet Series
TITANTEST™ Dry Series
8.6.2.1 Cleaning the venting connection filter
1. Switch the device off at the mains plug.
2. Pull out the mains plug.
Fig. 24: Venting filter
a Venting connection
3. Loosen the cover on the left and right side using an 8 mm Allen wrench
(
see Chapter 8.4.2: "Removing the cover (M-Series only)", page 69).
Required material: Exhaust Fittings T10019
Venting filter
Belüftungsfilter