LACO ORIGINAL OPERATING INSTRUCTIONS TITANTEST™
84
© 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807
8.8 Fuses
8.8.1 Replacing the fuses
1. Switch the device off at the mains plug.
2. Disconnect the device from the power supply and pull out the mains
plug.
3. Pry the cover of the fuse holder and fold it to the side.
Fig. 27: Replace the mains fuses (1)
4. Remove the two fuse holders and replace the defective fuses (10.0 AT,
250
V, Ø5 x 20 mm).
Fig. 28: Replace the mains fuses (2)
5. Lock the fuse holders back into place.
6. Close the cover.
7. Re-connect the mains plug.
Life threatening hazard from electric shock
Incorrectly secured products may be life threatening.
► Only use fuses with the prescribed values.