ECO GOLD
25/08/2011/ YACE0088 Operation 39
Caution !
Bath overflow due to thermal expansion or
immersion of objects
Burns, scalds, frostbite
• Take the volume of external consumers into account.
• Take into account the increase in volume with a rise
in temperature.
Caution !
Hot vapor formation / discharge of boiling cooling-water
on the cooling coil
Burns, scalds
• Filling of cooling coil with cooling water only admissi-
ble up to T
max
of 100 °C!
Notice
Inadmissible operating temperatures;
temperature difference between outflow and product
too large
Property damage (consumers, external components)
• Note that an externally controlled bath temperature,
especially during a transient response, may differ
substantially from the set-point temperature.
• Note the various limitation options (Tih, Til, T
max
, cor-
rection limitation).
• Set the overtemperature switch-off point T
max
accord-
ing to the heat transfer liquid (at least 25 K below the
fire point/boiling point).