ECO GOLD
8 Safety 25/08/2011/ YACE0088
1.3 Special safety information
The use of the thermostat is only admissible under the following conditions:
The siting surface must be impervious, flat, non-slip and non-combustible. Do not position the ther-
mostat at the edge of the bench or table.
Keep to the specified wall spacing (Ì 6.1).
Protect the th
ermostat from dripping or condensing water.
Do not store any liquids or combustible objects above the device.
Do not work with flammable liquids in the direct vicinity of the device.
Only connect the device to an earthed mains socket which is freely accessible.
At higher operating temperatures parts of the bath cover can take on temperatures of over 70 °C.
There is a danger of burns.
Only use suitable hoses. (Ì 6.4)
Ensu
re that the hoses are not kinked during operation.
Check the hoses at certain inspection intervals (Ì 8.3.2) for material fatigue.
Ho
ses with hot heat transfer liquid and other hot parts must not come into contact with the mains
cable.
When using the thermostat as a circulation thermostat, hot liquid can escape due to hose fracture
and become a danger to personnel and materials.
Toxic vapors may be generated depending on the heat transfer liquid used and the operating mode.
- Ensure sufficient extraction of the vapors.
- Use the bath cover.
Carefully mount the immersion thermostat on the bath vessel.
Only use bath vessels which are suitable for the intended operating temperatures.
When filling, set the overtemperature switch-off point according to the heat transfer liquid used.
When changing the heat transfer liquid from water to other liquids for temperatures above 100 °C,
carefully remove all residues of water including from the hoses and consumers, otherwise there is a
risk of scalding due to delay in boiling.
To do this also remove the blank plugs on the pump inputs and outputs and blow them out with com-
pressed air.
Use the cooling coil with cooling water only at operating temperatures below
100 °C. At higher tem-
peratures there is danger of hot vapors forming.
Have repairs carried out only by specialists.
Keep to all the service and maintenance intervals according to VDI 3033. (Ì 8.3.2)
Tak
e note of all safety labels.
Applicable to water-cooled devices:
Secure the return hose of the water cooling in the discharge area in order to prevent the hose sliding
off uncontrollably, also during pressure surges.
Secure the return hose of the water cooling in the discharge area so that it is not possible of hot cool-
ing water to splash out.
Avoid kinking or crushing the return hose of the water cooling. Excessive pressure can cause the
cooling water hoses to tear and hot water to escape.
To avoid damage due to a leak in the cooling water system we recommend the use of a water leak-
age sensor with water cut-off.