ECO GOLD
82 Appendix with settings 25/08/2011/ YACE0088
A.2 Setting the volume of the acoustic signals
The ECO GOLD thermostats sound alarms and faults as a two-tone acoustic signal. Warnings a signaled
as a continuous tone,
− Access to the main menu level is obtained by pressing the
enter key
.
− Selection and confirmation of Æ Setup Æ Basic setup Æ
Sounds.
− Choose Alarm, Warn or Error.
The adjacent menu window appears.
− The volume is selected with
or . The selected level is
immediately active without confirmation. In this example the
volume is medium.
− By pressing
(ESC), or you are returned to the menu level without any change.
A.3 Setting the chiller
The chiller of the cooling thermostats is operated in the "automatic" operating mode as standard. Here,
the cooling unit switches on or off automatically depending on the temperature and operating status.
However, you can also switch the cooling unit on or off manually.
− Access to the main menu level is obtained by pressing the
enter key
.
− Selection and confirmation of Æ Setup Æ Cooling.
The adjacent menu window appears.
(As well as Cooling the current operating mode is stated. This
is "off", "on" or "automatic".)
− Select Cooling with .
− With
or and you select and confirm the operating
status "off", "on" or "automatic".
− In the menu the set operating status is displayed by
a tick .
− By pressing
or (ESC) you are returned to the menu level without any change.
Note: When the cooling unit is switched off, it can take up to two minutes before it switches on again.
ESC
סּ OK
Standby
Warn
loud
medium
low
off
ESC
סּ OK
Standby
Setup
Pump level ►
Control ►
Cooling autom. ►
Limits ►
Basic setup ►
Calibration ►
ESC
סּ OK
Standby
Cooler
off
on .
automatic