ECO GOLD
25/08/2011/ YACE0088 Maintenance 55
8.2.3 Software version
SW version
confirm with
.
Under the menu point SW version the appropriate software
versions are displayed, depending on the device type and con-
nected modules.
8.2.4 Displaying and changing the device type
Type
confirm with
.
The device type without the suffix "G" (GOLD) is shown in the menu.
You can change the device type .
Note:
With a change of device type parameters are re-initialized and control parameters adapted by the
user are lost! Therefore, the type change has a three second delay on the key depression.
The overtemperature switch-off point T
max
is automatically adapted to the device type, i.e. with the ECO
GOLD thermostat with a stainless steel bath T
max
= 202 °C, for the ECO GOLD thermostat with transpar-
ent bath T
max
= 102 °C.
You must now manually re-enter T
max
(Ì 7.4.1), because otherwise the device enters the error status
(error message in ECO GOLD "T max diff. Ctrl-Safety").
8.2.5 Displaying serial numbers
Serial numbers
confirm with
.
Under the Serial numbers menu point the serial numbers of
Control and Safety are displayed. Provided they are available,
the serial numbers of connected modules are also displayed.
ESC
OK
Standby
SW version
Control 1.31.00
Safety 1.31
Cool 1.27
Ext Pt 1.21
Analog 3.01
Serial 3.03
ESC
OK
Standby
Serial numbers
Control
LCK1913-11-0002
Cool
KTExy-08-9876
Ext Pt

Table of Contents