X-Stream Operator’s Manual 
WM-OM-E Rev I  215 
Fx tab, for example  . The setup dialog for that Fx position appears. 
6. Touch the Single function button 
  if you want to perform just one math function on 
the trace, or touch the Dual function button 
  to perform math on math. 
7. Touch the Graph button, then touch inside the Graph with field to select a graph mode. 
The Graph modes are as follows: 
 
 
Histogram of the values of a parameter 
 
Track of the values of a parameter 
 
Trend of the values of a parameter 
Resampling To Deskew 
Deskew whenever you need to compensate for different lengths of cables, probes, or anything else 
that might cause timing mismatches between signals. Resample a signal on one channel and 
adjust it in time relative to a signal on another channel. 
To Resample 
1.  In the menu bar, touch Math, then Math Setup... in the drop-down menu. 
2.  Touch a math function trace tab F1 through Fx The number of math traces available 
depends on the software options loaded on your scope. See Specifications.. 
3. Touch the single function button. 
4.  Touch inside the Source1 field and select a source: channel, math trace, memory location. 
5. Touch inside the Operator1 field and select Deskew from the Functions category. 
6.  In the dialog on the right, touch the Deskew tab. 
7. Touch inside the Delay by data entry field and type in a time value, using the pop-up 
keypad.