. By selecting All, the test will be true only if every waveform falls within the limit 
that you will set. By selecting Any, the test will be true if just one waveform falls within the 
limit. 
7. Touch inside the Condition field in the "ParamCompare" mini-dialog and select a math 
operator from the pop-up menu: 
 
. 
8. Touch inside the Limit field and enter a value, using the pop-up numeric keypad. This 
value takes the dimension of the parameter that you are testing. For example, if you are 
testing a time parameter, the unit is seconds. 
9.  If you chose either WithinDeltaPct 
 or WithinDeltaAbs   from the Condition 
menu, touch inside the Delta field and enter a value.  
Mask Testing 
1. Touch Analysis in the menu bar, then Pass/Fail Setup... in the drop-down menu. 
2. Touch a Qx tab; a setup dialog for that position will open. 
3. Touch inside the Source1 field and select a source from the pop-up menu. 
4. Touch inside the Condition field in the main dialog and select Mask Test 
 . 
5.  From the "Test" mini-dialog, make a selection in the Test is True when group of buttons.