4D-Nucleofector
®
Manual
|
29
1. Catridge inseted in slot
2. Shot venting tube with filter attached to the male Luer
connector on the front plate
3. Tubes inseted in pinch valves (press black round
surface of the valves while pulling tubes fully into the
valves)
4. T connector inseted into T connector holder. Two
larger diameter tubes completely inseted into liquid
sensors
5. Upper tube (for cell suspension!) inseted into the
upper pump. Flap closed.
6. Optional - when feeding substrate separately - lower
tube inseted into lower pump. Flap closed.
7. Holder for cell suspension reservoir placed on a
magnetic stirring platform (stirring speed to approx.
300 rpm)
8. All reservoir-tube connections established to
a. Input reservoir for cell suspension
b. Input reservoir for substrate (optional)
c. Output reservoir
Figure 2.38: Setup Checklist for LV Nucleocuvette
®
Catridge. The picture
reflects the GMP-grade TheraPEAK
®
Consumables, research-grade versions
may slightly differ but the mounting procedure stays the same.
2
8c8b
4
6
5
7
8a
3
1
2
Return to TOC