EasyManua.ls Logo

Lucent Technologies Galaxy 4848/100 - Troubleshooting Procedure; Cabinet Alarm; Figure 12-1: Location of Cabinet Alarm Light and Controller; Purpose

Lucent Technologies Galaxy 4848/100
140 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Lucent Technologies Galaxy Power System 4848/100
12 - 2 Troubleshooting Preparations Issue 5 June 2000
Troubleshooting Procedure
Purpose The troubleshooting procedure described below is used when a trouble
condition has been identified and a technician has been dispatched to the
system location as a first and fundamental step in diagnosing and
correcting the problem.
For all trouble conditions, proceed as follows:
Cabinet Alarm 1. Locate the system controller. The controller is typically located in
the cabinet identified as BAY ONE. Because a trouble condition
exists, the red alarm on the top of the cabinet will be illuminated.
See Figure 12-1.).
Figure 12-1: Location of Cabinet Alarm Light and Controller
Galaxy SC
Controller
Cabinet Alarm Light
Galaxy
ESD
WRISTSTRAP
GROUNDING
POINT
Galaxy
Millennium
Controller
Cabinet Alarm Light

Table of Contents

Related product manuals