EasyManua.ls Logo

Lutz BE10 - Setting the Display Options; Calibrate; Calibration with a Comparative Measurement in Manual Mode

Default Icon
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
35
0698-651 Operating unit - 04/2021
en
9.6.1 Setting the display options
9.7 Calibrate
Theaccuracyoftheowmeterisadjustedtotherespectiveapplicationwiththecalibration.Thevolume
identiedbytheowmeteriscomparedwithareference(measurementvessel,scales,etc.).Thecalibra-
tion factor is changed accordingly in the case of a deviation.
Theowmetercanbecalibratedintwoways:
In calibration with a comparative measurement, the volume found by the
owmeteriscomparedwithareferenceornormvalue(e.g.meas.vessel).
Onceyouhaveenteredtheref.value,theowmetercalculatesthenewcal.
factorfromthedierencebetweenthetwovalues.
In the case of manual input, the calibration factor is entered directly via the
keypad.
9.7.1 Calibration with a comparative measurement in manual mode
Now add the required
amountofuidtoa
measurement vessel.
Current setting
Operating mode
Display options
Calibrate
Total volume
Menu 1/5
?
?
Display opt. 1/2
?
?
OK
Cancel
Partial volume/
Total volume
Partial volume/
Flow rate
Flow rate/
Total volume
Display opt. 1/2
?
?
OK
Cancel
Partial volume/
Total volume
Partial volume/
Flow rate
Flow rate/
Total volume
Menu
?
RESET
MAN
Calibrate
?
?
Cancel
Comparative
measurement
Manual input
Menu
?
RESET
MAN
Current setting
Operating mode
Display options
Calibrate
Total volume
Menu 1/5
?
?
Calibrate
?
?
Cancel
Comparative
measurement
Manual input
Cal. process
?
?
OK
Cancel
Press "OK" and
add the required
amount of fluid
to a vessel.
Then press
Cal. process
?
?
OK
Cancel
"OK" again and
enter the read
value from the
vessel.
Menu
?
OK
MAN

Table of Contents