Series F Test Frames + IntelliMESUR
®
Software User’s Guide
36
TOC
Distance Results
Load Averaging
Load Averaging dynamically captures data points and averages the result over a distance or
period of time. This requirement is common in peel testing and coefficient of friction testing
applications.
Enable the result, and any required statistics, then configure the following settings:
Trigger Force
The minimum force required to start the averaging sequence. Select compression or
tension. To start averaging immediately, set this value to zero.
Post-Trigger Delay
An initial delay may be configured, for example to omit the initial force spike at the
beginning of a peel test. To start averaging immediately after the trigger force, set this
value to zero.
Stop Condition
Stop averaging when the specified distance is reached or time period elapses.
The following graph is a representative example of a time-based averaging test with post-trigger
delay:
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Mark-10 F Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mark-10 F Series and is the answer not in the manual?

Mark-10 F Series Specifications

General IconGeneral
BrandMark-10
ModelF Series
CategoryTest Equipment
LanguageEnglish

Related product manuals