Series F Test Frames + IntelliMESUR
®
Software User’s Guide
Distance Results:
9.2 Step Types
Multi-step tests build upon the same functions as in Basic Tests, while separating out datum, prompting,
zero, and go-home functions. Preload is not integrated into steps as in Basic Tests. Instead, a preload
must be manually configured as a separate Load Limit step, followed by a Zero step.
Available steps are as follows:
9.2.1 Load, Distance, and Break
Each of these steps may be set up in the same manner described in Basic Test setup.
9.2.2 Hold
Hold a load (as described in Basic Test setup) or hold a position. Specify the hold period in hours,
minutes, and seconds.
To include the data points received during the Hold step in the Results calculations, enable Display
Results.