MI 3102(H) BT EurotestXE Technical specifications
93
10.5 Fault loop impedance and prospective fault current
10.5.1 No disconnecting device or FUSE selected
Fault loop impedance
Measuring range according to EN 61557 is 0.25 9.99k.
(5 % of reading + 5 digits)
Prospective fault current (calculated value)
Consider accuracy of fault
loop resistance measurement
The accuracy is valid if mains voltage is stabile during the measurement.
Test current (at 230 V) ........................... 6.5 A (10 ms)
Nominal voltage range ........................... 93 V 134 V (45 Hz 65 Hz)
185 V 266 V (45 Hz 65 Hz)
10.5.2 RCD selected
Fault loop impedance
Measuring range according to EN 61557 is 0.46 9.99 k.
(5 % of reading + 10 digits)
Accuracy may be impaired in case of heavy noise on mains voltage.
Prospective fault current (calculated value)
Consider accuracy of fault
loop resistance measurement
Nominal voltage range ........................... 93 V 134 V (45 Hz 65 Hz)
185 V 266 V (45 Hz 65 Hz)
No trip out of RCD.