EasyManua.ls Logo

MICRO-EPSILON thermoMETER CT Series - 7.2 Power Supply

MICRO-EPSILON thermoMETER CT Series
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 34
Electrical Installation
thermoMETER CT
7.1.1.2 CTM-1, CTM-2, CTM-3 Models
PIN Designation
+8 ... 36 VDC Power supply
GND Ground (0 V) of power supply
GND Ground (0 V) of internal in- and outputs
AL2 Alarm 2 (Open collector output)
OUT-TC Analog output thermocouple (J or K)
OUT-mV/mA Analog output object temperature (mV or mA)
F1-F3 Functional inputs
GND Ground (0 V)
3V SW 3 VDC, switchable, for laser sighting tool
GND Ground (0 V) for laser sighting tool
BROWN Temperature probe sensor (NTC)
WHITE Sensor ground
GREEN Sensor power Fig. 8 Opened controller
(CTM-1, CTM-2, CTM-3) with terminal
connections
YELLOW Detector signal
7.2 Power Supply
Please use a power supply unit with an output voltage of 8 - 36 VDC which can supply 100 mA. The ripple
should be max. 200 mV.
Please do never connect a supply voltage to the analog outputs.
> Destruction of the output
The thermoMETER CT is not a 2-wire sensor!

Table of Contents