Appendix | Accessories
thermoMETER CT
A 1.4 Further Accessories
A 1.4.1 Right Angle Mirror
The right angle mirror [TM-RAM-CT] for optics with D:S ≥ 10:1 enables measurements with 90 ° angle to sen-
sor axis.
Fig. 43 Right angle mirror [TM-RAM-CT]
The mirror has a reflection of 96 % in combination with a CT-SF22 and CT-SF15 and 88 % with a CTF-SF15.
If the mirror is used this value has to be multiplied by the emissivity value of the measurement object.
Example: CT-SF22 and object with emissivity = 0.85
0.85 x 0.96 = 0.816
Thus the emissivity in the CT-SF22 has to be set to the resulting value of 0.816.