EasyManua.ls Logo

MICRO-EPSILON thermoMETER CT Series - A 1.4 Further Accessories

MICRO-EPSILON thermoMETER CT Series
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 95
Appendix | Accessories
thermoMETER CT
A 1.4 Further Accessories
A 1.4.1 Right Angle Mirror
The right angle mirror [TM-RAM-CT] for optics with D:S ≥ 10:1 enables measurements with 90 ° angle to sen-
sor axis.
Fig. 43 Right angle mirror [TM-RAM-CT]
The mirror has a reflection of 96 % in combination with a CT-SF22 and CT-SF15 and 88 % with a CTF-SF15.
If the mirror is used this value has to be multiplied by the emissivity value of the measurement object.
Example: CT-SF22 and object with emissivity = 0.85
0.85 x 0.96 = 0.816
Thus the emissivity in the CT-SF22 has to be set to the resulting value of 0.816.

Table of Contents

Related product manuals