Page 71
Operating
thermoMETER CT
Display Mode [Sample] Adjustment range
P---- Signal output peak hold [inactive] P---- = inactive / 0.1 ... 999.9 s / P ∞ ∞ ∞ ∞ =
infinite
V---- Signal output valley hold [inactive] V---- = inactive / 0.1 ... 999.9 s / V ∞ ∞ ∞ ∞ =
infinite
u 0.0 Lower limit temperature range [0 °C] Depending on model / inactive at TCJ and TCK
output
n 500.0 Upper limit temperature range [500 °C] Depending on model / inactive at TCJ- and TCK-
output
[ 0.00 Lower limit signal output [0 V] According to the range of the selected output
] 5.00 Upper limit signal output [5 V] According to the range of the selected output
U °C Temperature unit [°C] °C/°F
/ 30.0 Lower alarm limit [30 °C] Depending on model
// 100.0 Upper alarm limit [100 °C] Depending on model
XHEAD Ambient temperature compensation
[Sensor temperature]
XHEAD = sensor temperature/-40.0 ... 900.0 °C
(for CT-SF) as fixed value for compensation/
returning to XHEAD (sensor temperature) by
pressing
and together
M 01 Multidrop address [1] (only with RS485
interface)
RS422 mode
01 ... 32, RS422 (Press
button on M 01)
B 9.6 Baud rate in kBaud [9.6] 9.6/19.2/38.4/57.6/115.2 kBaud
S ON Laser eyepiece (3 VDC-switches to the con-
nection pin „3 VSW“)
ON/OFF; This menu item appears on the models
CTM-1, CTM-2, CTM-3 on first position.