Milestone Srl MA213-003 – DMA-80 EVO – User Manual
140
CFR Part 11 is activated and one of the methods
is not signed. Please replace such method with a
new signed one.
Sample boats can be measured only once. The
software prevents possible double occupancy. If
needed, the position can be changed after the
start by the sample marked with the hourglass.
Add a valid calibration under measurement/link. At
least one cuvette must be analysed.
Alternative: measure calibration points.
When starting a measurement for calibration or
quality control (reference), the reference value
must be introduced.
In case of blank value, input the value 0.
A maximum of 99 points can be considered for the
calibration. Please create a new calibration or
delete old points.
Deactivation is not enough.
The heating coils are not properly heating or the
temperature is out of standard working range.
There could be problems in the heating element
but also temporary interruption of communication
or the lid of the unit has been opened and the
230V circuit cut.
Setting of the peak monitoring have noted a too
early or too late growth. This can occur i.e. in case
of leak.