Milestone Srl MA213-003 – DMA-80 EVO – User Manual
141
The pneumatic actuator are not in their starting
position. From system page you can move them
With automatic storage, after the creation of a
new sample list, this must be saved.
The detector signal has not reached its starting
value before peak detection. This could be caused
by excessive sample reaction, contamination of
the cuvette or electrical problem of
spectrophotometer components.
In case of excessive reaction of the sample, try to
reduce the amount and use a longer and gentler
analysis method.
- Normal sample reaction
- Excessive sample reaction
The detector voltage is not able to return to its
starting value