Milestone Srl MA213-003 – DMA-80 EVO – User Manual
65
If selected in grey colour, it means that the point is resulted out of the
absorbance limit of the cell but the the curve considers it in its track and
calculation.
You should uncheck these points.
This normally happens in cell 0 since it has a dedicated detector and/or in cell 1
to an existing calibration, when you change the limit of absorbance of the cell
(Settings/Analysis page. Please see chapter 5.17.1 )
Ng values: Values in “ng” of the calibration points
Abs values: Values of the absorbance (height of the peak) of the calibration
points
Deviation values from the calibration curve:
∆A: is the deviation of the absolute absorbance
∆A%: is the percentage deviation of the measured values
∆Hg[ng]: is the deviation in amount of mercury
∆Hg[%]:is the percentage deviation of amount of mercury
Example: