Milestone Srl MA213-003 – DMA-80 EVO – User Manual
83
With all data records a weight must always be entered, since otherwise the
measurement cannot be started.
The sample weight is indicated in “g”, “mg” or “µg” according to settings.
The sample weight is entered either manually by the keyboard/touch screen or
taken directly from an electronic balance through the interface cable.
Direct download of weighing: in most of balance models, the
weight is constantly updated on the terminal. In the sample table,
touch the “Weight” icon to download the value into the active cell.
For other balances: touching the print button of the balance (see
appropriate manual of the balance) the weight is transferred on
the terminal and into the active cell.
Automatic taring: In most balance models the balance can be
tared directly from the termina
Stop: the balance screen is closed without loading the weight.
Automatic multiple measurement (n) (only available in sub-section “Sample”):
Here multiple measurements can be activated. The basic requirement is that a
new row is added and you work under automatic mode. In this last line (and only
there), the desired number can be input in the column “n”.
In this way, it’s possible to add several lines. If in the starting line a sample
amount is entered, this is applied to all measurements. Even method and name
are taken. The first line is kept, since it is used for the results.
At the end of the analysis, in the first line of multiple measurements, the average
result is represented with related standard deviation. Single values can be
deactivated, while the sum sign is removed.
Additional points to be noted:
- Only in the first line marked by “M” the calibration factor is displayed.
This applies also to all measurements.