A5/A3â„¢ Service Manual 046-001141-00 4 - 73
Calibration System Calibration
FIGURE 4-92
3. Disconnect the hose connected to the pipeline assembly and attach it to the pressure meter
(For O2 hose #59; for AIR hose #54; for N2O hose #62).
FIGURE 4-93
4. Set the fresh gas flow of the gas being adjusted to 1 L/min.
5. Remove the self-locking acorn nut at the head of the regulator. Install the cylinder.