7-6
7.4 Upgrading Software
You can upgrade the software with the System Update Tool (PN: 110-000493-00) through
network.. This tool can directly run on a PC. Through network or by connecting the patient
monitor to a PC via a crossover network cable, you can upgrade the following programs:
Software PN Description Remark
6800-30-50770 BIOS program
Bootstrap program for
monitor main control unit,
implementing hardware
drive and initialization
For CE
Boot
program
110-001594-00 BIOS program For FDA
6800-30-50701 System software FDA
System
software
110-000414-00
System software CE
6800-30-50702 Language library FDA
Multilingual
library
110-000421-00
Language library CE
6800-30-50767 Icon library FDA
110-000422-00 Icon library CE
110-000807-00 Startup Logo CE
BMP
resource file
6802-30-66876 Startup Logo FDA
110-000815-00
FPGA configuration chip
software for mother board
CE
FPGA
program
6802-30-66895
FPGA configuration chip
software for mother board
(MDDR)
FDA
6800-30-50872
Audio data software for mother
board
FDA
Audio data
file
110-001423-00
Audio data software for mother
board
CE
Module rack
program
6800-30-50863
FPGA configuration chip
software for Module rack NIOS
board
/
MPM
module
0010-30-43083
BIOS program for parameter
module(BF531)
The System Update Tool
for MPM