4.4 Software Version Check 
Some troubleshooting tasks may require you to identify the configuration and software 
version of your patient monitor. 
1.  To view information on the system configuration and system software version, Select 
[Main Menu]→[Maintenance>>]→[Software Version>>]. 
2.  You can also view the information on system software version and module software 
version by selecting [Main Menu]→[Maintenance>>]→[Factory Maintenance>>]→ 
enter the required password  →[Software Version>>]. 
 
 
4.5 Technical Alarm Check 
Before troubleshooting the patient monitor, check for technical alarm message. If an alarm 
message is presented, eliminate the technical alarm first. For detailed information on 
technical alarm message, possible cause and corrective action, refer to the patient monitor’s 
Operation Manual. 
 
 
4.6 Troubleshooting Guide 
4.6.1 Power On/Off Failures 
Symptoms   Possible Cause  Corrective Action 
AC mains not connected 
or battery too low 
Check that AC mains is properly connected or 
battery capacity is sufficient. 
Power supply protection Refer to 4.6.9 Power Supply Failures . 
Cables defective or 
poorly connected 
1. Check that the cables from power switch & LED 
board to button board, button board to main 
board, and power module to main board are 
correctly connected.   
2. Check that cables and connectors are not 
damaged. 
Power switch & LED 
board defective 
Replace the power switch & LED board. 
Power module defective  Replace the power module. 
The patient 
monitor fails to 
start. AC LED 
or battery LED 
does not light   
Mother board Defective  Replace the mother board.