Mindray BeneView T5 - Module Performance Tests; ECG Tests and Calibration

Mindray BeneView T5
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-15
Block the gas inlet of the module or watertrap. The sidestream
CO
2
flowrate is slower than 10ml/min and an alarm of CO
2
Filterline Err is given. It indicates that there is no leakage.
The displayed CO
2
value is within 6±0.05.
Microstream CO
2
test
Block the gas inlet of the module or watertrap. An alarm of CO
2
Filterline Err is given. It indicates that there is no leakage.
The displayed CO
2
value is within 6±0.05.
AG test
When AG flowrate is slower than 10ml/min, an alarm of AG
Airway Occluded is given. It indicates that there is no leakage.
The fan inside the AG module works properly.
The measurement accuracy of CO
2
, N
2
O, O
2
and AA (AA
represents an anaesthetic agent) meets the product specifications
in the Operator’s Manual.