3-17 
 
ECG Calibration 
Tool required: 
  Vernier caliper 
 
Follow this procedure to perform a calibration: 
1.  Select the ECG parameter window or waveform area→ [Filter]→ [Diagnostic]. 
2. Select [Main Menu]→ [Maintenance>>]. 
3. Select [Calibrate ECG]. A square wave appears on the screen and the message [ECG 
Calibrating] is displayed. 
4.  Compare the amplitude of the square wave with the wave scale. The difference should 
be within 5%. 
5.  After completing the calibration, select [Stop Calibrating ECG]. 
 
If necessary, you can print out the square wave and wave scale through the recorder and then 
measure the difference. 
 
 
3.4.2 Resp Performance Test 
Tool required: 
  Fluke Medsim 300B patient simulator recommended 
 
Follow this procedure to perform the test: 
1.  Connect the patient simulator to the module using a non ESU-proof cable and set lead II 
as the respiration lead. 
2.  Configure the simulator as follows: lead II as the respiration lead, base impedance line 
as 1500 Ω; delta impedance as 0.5 Ω, respiration rate as 40 rpm. 
3.  Check the Resp wave is displayed without any distortion and the displayed Resp value is 
within 40 ± 2 rpm.