3.11.3 Installation and Test Report 
Basic Information 
Hospital 
 
Department 
 
Serial number of ID 
adapter 
 
Name of the external device 
 
ID of the external device 
 
Type of serial port adapting 
cable 
 
Software version and other information of 
the external device 
 
Checking the connection 
Test Result
(Ye s/ No ) 
Can the patient monitor and the external device be assembled together using 
designated accessories? 
 
Does the green indicator of corresponding port on the BeneLink module illuminate 
while the other indicators are not? 
 
Are there numerics or characters displayed on the [Devices Integrated] screen of the 
patient monitor? 
 
Are the device type and parameter values displayed correctly on the [Devices 
Integrated] screen of the patient monitor when the external device just enters working 
mode? 
 
Is the ventilation mode correctly displayed on the patient monitor? Does it change 
correctly when the ventilation mode on the external device is changed? 
 
Checking the parameters   
Parameters on the external 
device 
Value 
Parameters on the patient 
monitor 
Value 
PEEP  PEEP 
 
Pmean  Pmean 
 
VTe  VTe 
 
f ( RR )    f ( RR) 
 
Ppeak  Ppeak  
Checking the alarms 
Alarm 
Alarm messages displayed 
on the external device   
Alarm messages displayed in the 
alarm list of the patient monitor 
RR Too High     
Apnea    
Patient Disconnected