4-10
NOTE
z When the power module has a failure, it may cause problems to other components,
e.g. the monitor suddenly breaks down during start-up, as the power module may
have a power supply protection. In this case, troubleshoot the power module per
the procedure described in the table above.
z Components of the main unit, SMR and parameter modules are powered by the
power module. In the event that a component malfunctions, check if the operating
voltage is correct. Refer to 2 Theory of Operation for the operating voltage and
measurement points of each component.
4.6.10 Network Related Problems
Symptoms Possible Cause Corrective Action
Incorrect LAN cable
connection
Check LAN cable connection. LAN cable
shall not be longer than 50 m.
Frequent dropouts and
network disconnects
Incorrect IP address
configuration
Check for IP address conflict. Reconfigure IP
address.
Incorrect LAN cable
connection
Check LAN cable connection. LAN cable
shall not be longer than 50m.
Excessive requests for
viewing the patient
monitor at the same time
A patient monitor can only be viewed by 4
other patient monitors at the same time under
the View Others mode. The excessive view
requests system will be ignored.
Incorrect IP
configuration
Check for IP address conflict. Reconfigure IP
address.
The patient monitor is
connected to a LAN
but cannot view other
patients in the View
Others mode
USB_Hub board failure Replace the USB_Hub board.