3-24
5. Select [Respiratory Loop] in the [RM Setup] menu. Verify that the displayed TV is
within 500±50ml and RR is within 20±1rpm.
3.4.14 CCO/SvO
2
Tests
Interconnecting Function
Tools required:
CCO/SvO
2
cable (PN: 115-004083-00)
1. Connect and set the patient monitor and Vigilance monitor per the procedures in the
Operator’s Manual.
2. Set the Vigilance monitor to Demo mode.
3. Check that the CCO/SvO
2
numerics displayed on the patient monitor and Vigilance
monitor are consistent.
Output Performance
Tools required:
Oscillograph (Recommanded model: Agilent DSO5052A)
CCO/SvO
2
cable (PN: 115-004083-00)
1. Connect the signal output end of the connecting cables of the CCO/SvO
2
module to the
oscillograph.
2. Make the monitor to perform an ECG calibration. Check that the ECG waves displayed
on the oscillograph are consistent with the ECG calibration waves displayed on the
monitor screen.
3. Select [CCO Setup] [Signal Output Setup >>] and then select [Simulated High
Value] from the pop-up menu. Check that the amplitude of electrical level at the signal
output port of MAP, CVP and SpO
2
are 5±0.25V, 5±0.25V and 10±0.5V respectively.

Table of Contents

Other manuals for Mindray BeneView T5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T5 and is the answer not in the manual?

Mindray BeneView T5 Specifications

General IconGeneral
BrandMindray
ModelBeneView T5
CategoryHeart Rate Monitor
LanguageEnglish

Related product manuals