3-19 
 
3.4.6 IBP Tests 
IBP Performance Test 
Tool required: 
  Medsim300B patient simulator, MPS450, or other equivalent device 
  Dedicated IBP adapter cable (300B, P/N 00-002199-00) (use P/N 00-002198-00, if the 
simulator is MPS450) 
Follow this procedure to perform the test: 
1.  Connect the patient simulator with the pressure module. 
2.  Make the patient simulator outputs 0 to each IBP channel. 
3.  Press the Zero Key on the module to make a zero calibration. 
4.  Configure the patient simulator as P (static) = 200 mmHg. 
5.  The displayed value should be within 200 ± 2 mmHg. 
6.  If the value is outside of these tolerances, calibrate the pressure module. If the IBP 
module was calibrated with a dedicated reusable IBP sensor, check the calibration 
together with this IBP sensor. 
7.  Make that the patient simulator outputs 120/80 mmHg ART signals and 120/0 mmHg 
LV signals respectively to each IBP channel and check that the IBP wave is displayed 
correctly. 
8  Repeat the steps above for all the IBP channels. 
 
 
IBP Pressure Calibration 
Method 1: 
Tools required: 
  Medsim300B Patient simulator, MPS450, or other equivalent device 
  Dedicated IBP adapter cable (300B, P/N 00-002199-00) (use P/N 00-002198-00, if the 
simulator is MPS450) 
 
Follow this procedure to perform the test: 
1.  Connect the patient simulator to the pressure connector on the module. 
2.  Set the patient simulator to 0 pressure for the desired IBP channel. 
3.  Press the Zero Key on the module to make a zero calibration. 
4.  Configure the patient simulator as P (static) = 200 mmHg. 
5. Select [Main Menu]→ [Maintenance >>]→[User Maintenance >>] →[Cal. IBP 
Press. >>]. In the [Cal. IBP Press.] menu, set the calibration value to 200 mmHg.