3-25 
 
3.4.13 RM Test 
Tool required: 
  Gas source 
  Ventilator (calibrated) 
  Artificial lung 
  Pediatric/neonate flow sensor 
 
 
Follow this procedure to perform the test: 
1.  Connect the equipment as shown above. Make sure that the blue sensing tube on the 
flow sensor is connected with the artificial lung. 
2. Set [Patient Cat.] to [Adu]. In the [RM Setup] menu, select [Sensor Type] according 
to the used sensor and set [Ventilation Mode] to [Mechanical]. 
3. Enter the [RM Setup] menu and select [Calibrate >>]. Input the constant marked on 
the sensor and calibrate the flow sensor. 
4.  Configure the ventilator as follows: TV=500 ml, RR =20 rpm, I:E=1:2. 
5. Select [Respiratory Loop] in the [RM Setup] menu. Verify that the displayed TV is 
within 500±50ml and RR is within 20±1rpm. 
 
Ventilator 
Artificial 
lung 
Monitor
Flow sensor