External Device  Setup  Remark 
Ohmeda Avance/Aisys  Not required. 
The following information is for 
further reference: 
Baud Rate: 19200 bps 
Word Length: 7 bits 
Parity: odd 
Stop Bits: 1 
Ohmeda Aestiva 7100 
/7900 
Not required. 
The following information is for 
further reference: 
Baud Rate: 19200 bps 
Word Length: 7 bits 
Parity: odd 
Stop Bits: 1 
Drager Fabius GS/Plus/Tiro 
Protocol: Medibus 
Baud Rate: 9600 
Word Length: 7 bits 
Parity: even 
Stop Bits: 1 
/ 
Drager Primus 
Protocol: Medibus 
Baud Rate::9600 
Word Length: 8 bits 
Parity: even 
Stop Bits: 1 
/ 
TCM CombiM/TCM 
TOSCA 
Protocol: Monlink. 
The following information is for 
further reference: 
Baud Rate: 9600 bps 
Word Length: 8 bits 
Parity: even 
Stop Bits: 1 
TOF-Watch SX  Not required. 
The following information is for 
further reference: 
Baud Rate: 19200 bps 
Word Length: 8 bits 
Parity: none 
Stop Bits: 1