3-20 
6. Select the [Calibrate] button next to the desired IBP channel to start a calibration. 
7.   If the calibration is completed successfully, the message [Calibration Completed!] will 
be displayed. Otherwise, a corresponding message will be displayed. 
 
Method 2: 
Tools required: 
  Standard sphygmomanometer 
  Balloon pump 
  Tubing 
  T-shape connector 
 
To perform a calibration: 
1.  Connect the 3-way stopcock, the sphygmomanometer and the balloon pump through a 
T-shape connector, as shown below. 
2.  Zero the transducer. Then open the stopcock to the sphygmomanometer. 
3.  Press the Main menu button on the equipment’s front panel. Select [Maintenance>>]→ 
[User Maintenance >>]→ enter the required password→[Cal. IBP Press. >>]. Then 
configure IBP calibration value. 
4.  Inflate using the balloon pump until the reading of sphygmomanometer approximates 
the preset calibration value. 
 
 
 
5.  Adjust the calibration value in the [Cal. IBP Press.] menu until it is equal to the reading 
of sphygmomanometer 
6. Select the [Calibrate] button to start a calibration 
7.  The message [Calibration Completed!] is displayed after a successful calibration. If 
the calibration failed, the prompt [Calibration Failed!] will be displayed. 
 
 
Sphygmomanometer 
T-shape connector
3-way stopcoc
Pressure transducer
Pressure adapter cable 
IBP Module